Vendor DICOM Tag Reference

DICOM tags used by each MRI vendor for perfusion imaging parameters. These differences affect how osipy extracts metadata during data loading.

Standard DICOM Tags (All Vendors)

Tag Name Description
(0008,0070) Manufacturer Vendor identification
(0008,1090) ManufacturerModelName Scanner model
(0018,0080) RepetitionTime TR in ms
(0018,0081) EchoTime TE in ms
(0018,1314) FlipAngle Flip angle in degrees
(0018,9087) DiffusionBValue B-value for DWI
(0020,0013) InstanceNumber Volume/slice number
(0020,1041) SliceLocation Slice position

GE Medical Systems

Identification

  • Manufacturer: "GE MEDICAL SYSTEMS"
  • Private Groups: 0027, 0043, 0019

DCE-MRI Tags

Tag Name Notes
(0018,0080) RepetitionTime Standard TR
(0018,0081) EchoTime Standard TE
(0018,1314) FlipAngle Standard
(0019,10a2) NumberOfPhases Number of dynamic timepoints

DWI/IVIM Tags

Tag Name Notes
(0043,1039) B-value GE private b-value array
(0019,10bb) DiffusionBValue Alternative b-value location
(0019,10bc) DiffusionDirection Gradient direction

ASL Tags

GE ASL data is typically: - 3D spiral pCASL (product sequence) - Parameters in BIDS sidecar after conversion - M0Type: "Included" common (M0 in same timeseries)

Known Issues: - B-values may not populate standard DiffusionBValue tag - Must check private tags for complete diffusion info - ASL parameters often require BIDS sidecar

Siemens

Identification

  • Manufacturer: "SIEMENS"
  • Private Groups: 0019, 0021, 0029, 0051

CSA Headers

Siemens stores extended metadata in CSA (Cine Supplementary Attributes) headers: - (0029,1010) CSA Image Header Info - (0029,1020) CSA Series Header Info

The osipy Siemens parser extracts:

Extract metadata using Siemens parser

from osipy.common.io.vendors import SiemensParser

parser = SiemensParser()
metadata = parser.extract_metadata(ds)
# Accesses CSA headers for extended parameters

DCE-MRI Tags

Tag Name Notes
(0018,0080) RepetitionTime Standard TR
(0018,0081) EchoTime Standard TE
(0018,1314) FlipAngle Standard
(0051,100c) NumberOfImages May indicate timepoints
CSA Header MrPhoenixProtocol Detailed sequence parameters

DWI/IVIM Tags

Tag Name Notes
(0019,100c) DiffusionBValue Siemens private b-value
(0019,100d) DiffusionDirection Gradient direction (x)
(0019,100e) DiffusionDirection Gradient direction (y)
(0019,100f) DiffusionDirection Gradient direction (z)
CSA Header B_value B-value from CSA

ASL Tags

Siemens ASL sequences include: - WIP pCASL: 3D GRASE readout - PASL/FAIR: Bremen sequence (multi-PLD) - 2D EPI: Research sequences

Parameter Source Notes
ArterialSpinLabelingType BIDS sidecar PASL, PCASL
PostLabelingDelay BIDS sidecar May be array for multi-PLD
BolusCutOffFlag BIDS sidecar Q2TIPS for PASL
BolusCutOffDelayTime BIDS sidecar Q2TIPS timing
BolusCutOffTechnique BIDS sidecar "Q2TIPS"

Known Issues: - Multi-PLD data has array of PLDs (not single value) - CSA header parsing may fail on some sequences - Research sequences may have non-standard parameters

Philips

Identification

  • Manufacturer: "Philips Medical Systems" or "Philips"
  • Private Groups: 2001, 2005

Enhanced DICOM

Philips commonly uses Enhanced Multi-frame DICOM: - Single file contains all slices/timepoints - Nested sequences for per-frame metadata - Use dcm2niix for reliable conversion

DCE-MRI Tags

Tag Name Notes
(0018,0080) RepetitionTime Standard TR
(0018,0081) EchoTime Standard TE
(0018,1314) FlipAngle Standard
(2005,1030) PrivateRepetitionTime May differ from standard
(2001,1023) NumberOfDynamics Number of timepoints

DWI/IVIM Tags

Tag Name Notes
(2001,1003) DiffusionBValue Philips private b-value
(2001,1004) DiffusionDirection Gradient direction
(0018,9087) DiffusionBValue May also be populated

ASL Tags

Philips ASL typically uses 2D EPI pCASL:

Parameter Source Notes
ArterialSpinLabelingType BIDS sidecar Usually "PCASL"
PostLabelingDelay BIDS sidecar Single value typically
LabelingDuration BIDS sidecar Required for quantification
BackgroundSuppression BIDS sidecar Usually true

Known Issues: - DICOM headers often lack complete ASL parameters - Control/label order not specified in headers - Must document settings at acquisition time - LabelingDuration may be missing

Parameter Extraction Priority

The osipy metadata mapper uses this priority:

1. BIDS sidecar JSON
   └─ Highest confidence, standardized format

2. Vendor private tags
   └─ Parsed by vendor-specific parser

3. Standard DICOM tags
   └─ (0018,xxxx) series

4. User-provided values
   └─ Via function parameters

5. Modality defaults
   └─ Reasonable assumptions per modality

Verification Checklist

When loading multi-vendor data, verify:

DCE

  • [ ] TR is in expected range (typically 2-10 ms for fast DCE)
  • [ ] Flip angle(s) recorded
  • [ ] Number of timepoints matches expectation
  • [ ] Temporal resolution calculated correctly

DSC

  • [ ] TE is appropriate (typically 25-50 ms for gradient echo)
  • [ ] TR is in expected range (typically 1-2 s)
  • [ ] Number of baseline volumes known

ASL

  • [ ] Labeling type identified (PCASL, PASL, CASL)
  • [ ] PLD specified (or array for multi-PLD)
  • [ ] Labeling duration specified (for pCASL/CASL)
  • [ ] Control/label order known (from aslcontext.tsv)
  • [ ] M0 type identified (Separate, Included, Absent)

IVIM

  • [ ] B-values extracted correctly
  • [ ] B=0 acquisition present
  • [ ] Sufficient b-values for IVIM (≥4, ideally 6+)
  • [ ] Low b-values present (<200 s/mm²)

Testing with Real Data